Subscribers
Non-subscribers
- Purchase article PDF [£27 + taxes]
- Purchase article PDF member offer [£5 + taxes]
Free access
Technical Note
Lab Chip, 2008, 8, 1733 - 1737, DOI: 10.1039/b807037g
An active bubble trap and debubbler for microfluidic systems
Alison M. Skelley and Joel Voldman
We present a novel, fully integrated microfluidic bubble trap and debubbler. The 2-layer structure, based on a PDMS valve design, utilizes a featured membrane to stop bubble progression through the device. A pneumatic chamber directly above the trap is evacuated, and the bubble is pulled out through the gas-permeable PDMS membrane. Normal device operation, including continuous flow at atmospheric pressure, is maintained during the entire trapping and debubbling process. We present a range of trap sizes, from 2 to 10 mm diameter, and can trap and remove bubbles up to 25
L in under 3 h.