Issue 4, 2009

Isotropically etched radial micropore for cell concentration, immobilization, and picodroplet generation

Abstract

To enable several on-chip cell handling operations in a fused-silica substrate, small shallow micropores are radially embedded in larger deeper microchannels using an adaptation of single-level isotropic wet etching. By varying the distance between features on the photolithographic mask (mask distance), we can precisely control the overlap between two etch fronts and create a zero-thickness semi-elliptical micropore (e.g. 20 µm wide, 6 µm deep). Geometrical models derived from a hemispherical etch front show that micropore width and depth can be expressed as a function of mask distance and etch depth. These models are experimentally validated at different etch depths (25.03 and 29.78 µm) and for different configurations (point-to-point and point-to-edge). Good reproducibility confirms the validity of this approach to fabricate micropores with a desired size. To illustrate the wide range of cell handling operations enabled by micropores, we present three on-chip functionalities: continuous-flow particle concentration, immobilization of single cells, and picoliter droplet generation. (1) Using pressure differentials, particles are concentrated by removing the carrier fluid successively through a series of 44 shunts terminated by 31 µm wide, 5 µm deep micropores. Theoretical values for the concentration factor determined by a flow circuit model in conjunction with finite volume modeling are experimentally validated. (2) Flowing macrophages are individually trapped in 20 µm wide, 6 µm deep micropores by hydrodynamic confinement. The translocation of transcription factor NF-κB into the nucleus upon lipopolysaccharide stimulation is imaged by fluorescence microscopy. (3) Picoliter-sized droplets are generated at a 20 µm wide, 7 µm deep micropore T-junction in an oil stream for the encapsulation of individual E. coli bacteria cells.

Graphical abstract: Isotropically etched radial micropore for cell concentration, immobilization, and picodroplet generation

Supplementary files

Article information

Article type
Paper
Submitted
03 Oct 2008
Accepted
12 Dec 2008
First published
07 Jan 2009

Lab Chip, 2009,9, 507-515

Isotropically etched radial micropore for cell concentration, immobilization, and picodroplet generation

T. D. Perroud, R. J. Meagher, M. P. Kanouff, R. F. Renzi, M. Wu, A. K. Singh and K. D. Patel, Lab Chip, 2009, 9, 507 DOI: 10.1039/B817285D

To request permission to reproduce material from this article, please go to the Copyright Clearance Center request page.

If you are an author contributing to an RSC publication, you do not need to request permission provided correct acknowledgement is given.

If you are the author of this article, you do not need to request permission to reproduce figures and diagrams provided correct acknowledgement is given. If you want to reproduce the whole article in a third-party publication (excluding your thesis/dissertation for which permission is not required) please go to the Copyright Clearance Center request page.

Read more about how to correctly acknowledge RSC content.