JAAS, 2004, 19

Additions and corrections

Determination of Ir and Pt down to the sub-femtogram per gram level in polar ice by ICP-SFMS using preconcentration and a desolvation system

Paolo Gabrielli, Anita Varga, Carlo Barbante, Claude Boutron, Giulio Cozzi, Vania Gaspari, Frédéric Planchon, Warren Cairns, Sungmin Hong, Christophe Ferrari and Gabriele Capodaglio

JAAS, 2004, 19 (DOI: 10.1039/b316283d). Amendment published 24th May 2004

The published version of Table 2 of this paper is incorrect. Table 2 should be as follows: Table 2 Potential spectral interferences that could affect the determination of ultra trace levels of Ir and Pt in polar ice


Analyte
Potential interference
Resolution (m/m)
Isotope
Abundance (%)
Species
Abundancea
191Ir 37.30 175Lu16O 97.17 7667
173Yb18O 0.03 8222
174Yb17O 0.01 8449
193Ir 62.70 177Hf16O 18.51 7782
175Lu18O 0.19 8389
194Pt 32.90 178Hf16O 27.17 8063
177Hf17O 0.01 9547
176Hf18O 0.01 8776
176Yb18O 0.03 9260
176Lu18O 0.01 9312
195Pt 33.80 179Hf16O 13.60 8108
178Hf17O 0.01 8884
177Hf18O 0.04 8706
196Pt 25.20 180Hf16O 34.99 8348
180W16O 0.13 8408
180Ta16O 0.01 8692
178Hf18O 0.05 8875
179Hf17O 0.005 9803
196Hg 0.14 226550
a For polyatomic species calculated as the product of the natural abundances of each isotope divided by 100.  




The Royal Society of Chemistry apologises for these errors and any consequent inconvenience to authors and readers.


Back